Electromagnetic compatibility (EMC): Electromagnetic compatibility (EMC)
A-6 Dash 3000/4000/5000 2000966-542D
Recommended separation distances
The table below provides the recommended separation distances (in
meters) between portable and mobile RF communications equipment and
the Dash 3000/4000/5000.
The Dash 3000/4000/5000 is intended for use in the electromagnetic
environment on which radiated RF disturbances are controlled. The
customer or the user of the Dash 3000/4000/5000 can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the Dash 3000/4000/5000 as recommended below,
according to the maximum output power of the communications
equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equitation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance estimation
(Transmit power of 802.11a/b/g wireless communication device)
Power of the Source
(Watts)
Separation Distance (Feet) Separation Distance
(Meter)
0.01 0.75 0.23
0.1 2.39 0.73
17.542.3
10 23.94 7.3
100 75.44 23

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable