Preventive maintenance: Battery maintenance
2000966-542D Dash 3000/4000/5000 5-11
How to condition the battery
The battery can be conditioned by one of two methods:
Outside of a patient monitor by using the Cadex SMart Two+
charger.
Inside a patient monitor that is connected to an AC power source.
NOTE
To extend the life of the battery, GE recommends that you condition
the battery using the external Cadex SMart Two+ charger.
NOTE
The Cadex SMart Two+ charger is no longer available for purchase.
However, if your site has a Cadex SMart Two+ charger, you should
continue to use it to charge and condition the battery.
Conditioning the battery with a Cadex SMart Two+ charger
A conditioning cycle using the battery charger requires approximately
nine hours to complete. Complete the following steps to automatically
condition the battery.
1. Insert the battery into the battery charger.
2. Press the CONDITION button when one of the following conditions
occur:
While the RUN LED light is still flashing.
While the CONDITION LED is flashing.
3. Remove the battery from the battery charger when the RUN LED
lights. This completes the conditioning cycle.
Conditioning the battery inside a patient monitor
NOTE
Non-recommended (incompatible) batteries will not condition in the
Dash patient monitor.
WARNING
PATIENT RISK HAZARD — Never condition a battery
while the patient monitor is connected to a patient.
Serious injury or death could result.
Complete the following steps to manually condition the battery using the
patient monitor:
1. Remove all batteries from the patient monitor.
2. Insert one battery into Slot A of the patient monitor.
3. Disconnect the patient monitor from the AC power source.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals