Installation: Dash installation checkout procedure
3-8 Dash 3000/4000/5000 2000966-542D
Dash installation checkout procedure
GE recommends that qualified personnel shall perform the following
tests, and should record the results on the “Checklist” on page D-1
The Dash unit has been thoroughly tested before leaving manufacturing.
There is no need to re-test the parameters or operation of the unit upon
installation except for the following:
1. Electrical safety test. Refer to Electrical safety tests” on page 8-4.
2. Refer to “Patient monitor power-up tests” on page 8-21.
3. Battery check. Ensure the battery indicator displays on the patient
monitor without an error.
4. If the Dash is networked:
Connect Dash to the MC network (wired or wireless) with any
other GE bedside monitor.
Select View Other.
Select another bed.
Verify you can view another bed over the network.
5. If the Dash has TRAM-rac 2A housing connected via the Aux port:
Verify communication with a module in the TRAM-rac 2A
housing by confirming that the parameter for the ICG or SAM
module displays on the Dash.
6. If the Dash has a BISx module connected via the Aux port:
Verify communication with the BISx module by confirming that
the parameter for the module displays on the Dash.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals