Configuration: Advanced user procedures
4-24 Dash 3000/4000/5000 2000966-542D
Transfer monitor defaults
NOTE
Transfer Monitor Defaults feature is part of the CARESCAPE
Network software option and is available in software version 3 or
later. You must purchase and enable this software option before you
can transfer patient monitor defaults.
NOTE
To prevent the accidental erasure of the client patient monitor’s
wireless LAN defaults, make sure that the server and client patient
monitors’ wireless LAN defaults match before transferring patient
monitor defaults.
The patient monitor defaults set on one patient monitor (server) may be
transferred to another patient monitor (client), provided that the client
patient monitor shares the following:
the same software revision (version),
the same patient-monitor type (e.g., ADULT, NEO, or OR),
the same wireless LAN configuration,
the same software package (e.g., Basic, Cardiac, Cardiopulmonary),
and
the same country code (e.g., Default or France).
The following defaults transfer when using this feature:
all patient monitor defaults
wireless LAN configuration
custom default names
12 SL location
12 SL site

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable