Troubleshooting: Review errors
2000966-542D Dash 3000/4000/5000 6-15
NOTE
The patient monitor may be referred to as a display or scope in the
error code descriptions.
Severity
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Continue—the event/error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log entries
will have a severity of Continue.
Fatal—the event/error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a WARM
START.
Forced Restart—the system was restarted by a known condition
(internet address change, video test, etc).
Date and time
The date and time the event or problem occurred.
Error number
A sequential number that is used to identify each event/problem.
70F Date was changed from this patient monitor. This helps determine
how the system-wide date was altered.
710 Incompatible software was detected. If the main processor software
finds that the software operating on the communication software
incompatible, it turns off the communication (network) controller and
enters this data into the error log. When the patient monitor won’t
“talk” to the network, looking for this entry in the error log is one part of
the troubleshooting process.
Definition of error codes

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals