EasyManuals Logo

GE Dash 4000 Service Manual

GE Dash 4000
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Equipment overview: Theory of operation
2000966-542D Dash 3000/4000/5000 2-31
Real-time clock
The real-time clock incorporates an on-board quartz crystal. This feature
simplifies the design and eliminates adjustments. The time of day on a
standalone patient monitor is maintained to an accuracy of 15 seconds in
24 hours to support 24 hour trend data. The RTC device is one of the
devices on the I
2
C bus.
Audio subsystem
Audio tones are generated using a tone generator, an audio amplifier,
and an 8, 2.5-in. speaker. The tone generator has built in D/A
converters and a mixer to generate the dual frequency tones.
Frequencies ranging from 150 Hz to 2800 Hz are produced.
Video subsystem
The microprocessor contains a video controller that supports real-time
and non-real-time waveform drawing, menu drawing, and parameter
display.
Display brightness is controlled by a 50 k digital potentiometer. The
potentiometer is interfaced to the brightness control input of an inverter.
Analog outputs
Two analog output channels support ECG and BP. The pace pulse is
generated on the main processor board and inserted into the analog out
ECG signal. The digital-to-analog conversion for both ECG and blood
pressure output signals are performed on the main processing board via
a two channel 12-bit serial DAC. Cal data is stored on the processor/
power management PCB.
Defib Sync
The QRS complex of ECG data acquired from the DAS generates the
marker-out signal. A software selectable pulse width and pulse
amplitude is provided in the Boot Loader Menu.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable