IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
7-16 Section 7-4 - Gathering Troubleshooting Data
PRELIMINARY
Section 7-7
Probes not Recognized Troubleshooting
7-7-1 Purpose of this Section
If the system does not recognize any of the connected probes. perform the following procedure.
7-7-2 Probes not Recognized Procedure
1.) Reseat the PSB, TCFE and TRX boards according to Venue service manual
2.) If the issue is not resolved, this may be a probe issue. In this case, perform the following steps for
each of the probes individually:
1.) disconnect all probes except one probe and restart the system
2.) If the issue persists, continue checking this probe on all the other ports to decide if it is a
probe issue or a PSB issue
3.) If the system still does not recognize any of the probes, replace PSB module (P/N S5726584)