8-66 Section 8-5 - Electronic Boards- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
Section 8-5
Electronic Boards- Replacement Procedures
8-5-1 BIB Board Replacement Procedure
8-5-1-1 Tools
Appropriate Phillips Screwdriver.
8-5-1-2 Time Required
15 min
8-5-1-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-5-1-4 BIB Board Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower, Left side eTower, and Right
Side eTower.
3.) Remove the Back End Module.
4.) Remove the supporting bracket using an appropriate Phillips screwdriver:
• Release two captive screws
Refer to Table 9-13 on page 9-13.
• Accessories - Replacement Procedures
• RS Probe Cover Removal Procedure
• Lower Front eTower Cover Removal Procedure
• Left Side eTower Cover Removal Procedure
• Right Side eTower Cover Removal Procedure
• BE Module Removal Procedure