D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
3-22 Section 3-5 - Completing the Setup
PRELIMINARY
Section 3-5
Completing the Setup
3-5-1 Purpose of this Section
This section describes how to complete the setup of the Venue™.
3-5-2 System Specifications
3-5-2-1 System Requirements Verification
• Verify that the site meets the requirements listed in Chapter 2.
(See: Facility Needs on page 2 - 7.)
• Verify that the specifications below do not conflict with any on-site conditions.
3-5-2-2 Physical Dimensions
3-5-2-3 Mass with Monitor and Peripherals
3-5-2-4 Acoustic Noise Level
Less than 55 dB(A) at 20 degrees Celsius, measured in the operators head position, 20 cm in front of
the keyboard’s right corner, at 1.30 m above the floor, and in a distance of 1 meter at all four sides, 1
meter above the floor.
3-5-3 Electrical Specifications
3-5-3-1 Verification of the Venue™ Voltage Setting
Verify that the mains voltage specified for the Venue™ is available on-site.
The voltage setting for the Venue™ is found on a label near the Mains Power Circuit Breaker on the
rear of the Venue™ .
3-5-3-2 Electrical Specifications for the Venue™
In the table below, the electrical specifications for Venue™ includes monitor and on board peripherals.
Table 3-12 Physical Dimensions of Venue™ with Monitor and Peripherals
in Transportation Position
Height Width Depth Unit
131 48 72.5 cm
51.57 18.89 28.54 Inches
Table 3-13 Mass of Venue™ with Cockpit (monitor), without Probes and Peripherals
Model Mass [KG] Mass [LBS]
Venue™ 63 139
Connecting a Venue™ to the wrong voltage level will most
likely destroy it.