IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-21
PRELIMINARY
7-8-6 No Auxiliary Display
Symptom: No Auxiliary Display (LEDs are off) when turning on the system
Suggested solution: Reset front end connections
1.) If fans are spinning:
a.) Reset cockpit cables
b.) Reset MPB
c.) Replace cockpit
2.) If fans are NOT spinning
a.) Turn off the switch located at the rear part of the system for 30 seconds and turn on again
b.) Reset cockpit cables
c.) Reset MPB
d.) Replace MPB
Figure 7-43 No auxiliary display