Chapter 8 - Replacement Procedures 8-33
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-3-5 Full Front End Installation
1.) Install the new FE assembly.
2.) Secure the Full FE Module to the FE Door with four captive screws.
NOTE: It is recommended to secure the upper left captive screw first.
3.) Mount the FE Door assembly on the scanner.
4.) Reconnect the following cables:
• PCIe cable
• MPB TO T-FEPS Cable and secure it to the CFE module and to the T-FEPS module with cable
clips.
• Fan cable
5.) Close the FE door and close the two fastening latches on the left side of the Front End door
assembly (Figure 8-19).
6.) Tighten the captive screw.
7.) Refit the following covers: Lower Front eTower, left side eTower, right side eTower and RS Probe
Cover.
8.) Install all accessories.
9.) Turn ON power to the system
10.)Perform Software Update Procedure to make sure the FW is updated
• Left Side eTower Cover Installation Procedure
• Right Side eTower Cover Installation Procedure
• Lower Front eTower Cover Installation Procedure
• RS Probe Cover Installation Procedure
• Accessories - Replacement Procedures
Perform the checks listed in Full Front End Replacement on page 8-217