8-38 Section 8-4 - System Modules- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
• Ground cables (x2))
5.) Remove the Gel Holder.
6.) Hold the cockpit (monitor) on its top, insert a screwdriver into the dedicated hole and press to
release the securing spring. The click is sounded, indicating the securing spring is released.
7.) Lift and remove the cockpit (monitor).
Figure 8-25 Cockpit Ground Cables Removal
Raise the Articulated Arm to its up position to prevent the unintentional injury to personnel from
the stored mechanical energy.
Figure 8-26 Cockpit (Monitor) Securing Screws
Insert screwdriver here
Spring released
Spring locked