SCANNER (OPTIONAL)
111
Run Scan
Overview
Scan operation
type
Scan
Measures all the specified channel ranges
at each trigger event. The timer settings
(page 106) apply to each scan.
Step
Measures a single channel in the specified
range at each trigger event. The timer
settings (page 106) apply to each channel.
Monitor
Continuously measures one channel.
Run Scan/Step
Activate
Scan/Step
1. Press the Shift key followed by
the ACV key (Scan) or DCV key
(Step).
ACV
SCAN
DCV
STEP
2. The STO indicator turns On. The Scan (Step) starts
running and the data is recorded. After running the
predefined count, the Scan (Step) stops running.
STO
Vm
SAUTODC
Retrigger/Restart
Scan
To run the Scan (Step) again, press
the Trig key. The previous data is
overwritten by the new Scan.
Abort Scan/Step
To abort Scan/Step or to go back to
the normal display, press the Shift
key followed by the ACV key (Scan)
or DCV key (Step) again.