BASIC MEASUREMENT
41
Range
RTD: -200°C ~ +600°C (sensor dependent)
Thermocouple: -210°C ~ +1820°C (sensor dependent)
Select Thermocouple Type
Background
The GDM-8261accepts thermocouple inputs and
calculates the temperature from the voltage difference of
two dissimilar metals. Thermocouple type and reference
junction temperature are also considered.
Parameter Thermocouple Range Resolution
E
-200 to +1000°C 0.002 °C
J
-210 to +1200˚C 0.002 °C
T
-200 to +400°C 0.002 °C
K
-200 to +1372˚C 0.002 °C
N
-200 to +1300˚C 0.003 °C
R
-50 to +1768˚C 0.01 °C
S
-50 to +1768˚C 0.01 °C
B
+350 to +1820˚C0.01 °C
1. Open sensor
selection menu
Press the Shift key, then the TEMP
(Sensor) key. The sensor selection
menu appears on the display.
TEMP
SENSOR
2. Select sensor
type
Press the Left and Right arrow keys
and select T-CUP (thermocouple).
3. Select sensor
Press the Down key twice. The
sensor selection menu appears on
the display.
www.GlobalTestSupply.com