EasyManua.ls Logo

GW Instek GDM-8261 - 2 W;4 W Resistance Measurement

GW Instek GDM-8261
195 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDM-8261 User Manual
32
2W/4W Resistance Measurement
Measurement
type
2-wire
Uses the standard V-COM ports.
Recommended for measuring resistances
larger than 1k.
4-wire
Compensates the test lead effect using the
4W compensation ports, in addition to the
standard V-COM ports.
Recommended for measuring sensitive
resistances smaller than 1k.
1. Activate
resistance
measurement
For 2-wire resistance measurement,
press the 2W/4W key once.
For 4-wire resistance measurement,
press the 2W/4W key twice.
2. 2W/4W
resistance mode
display appears
2W or 4W +
Indicates 2W or 4W Resistance mode
AUTO Indicates Automatic range selection
1K 2nd display shows the Resistance range
3. Connect the
test lead and
measure
Connect the test lead. For 2-wire resistance, use the (V)
and the COM port. For 4-wire resistance, use the (V)
and the COM port, plus the 4W sense, and LO port for
sensing. The display updates the reading.
2W connection
INPU
T
V
COM
4W connection
4W INPUT
SENSE
HI
LO
V
COM