GDM-8261 User Manual 
132 
Command Syntax 
The commands are partially compatible with IEEE488.2 (1992) and SCPI 
(1994) standard. Commands are NON-case sensitive.   
 
Example 
command 
CONF:VOLT:DC ,MAX
12345
10
 
1: Command Header 
2: Single space 
3: Parameter 1 
 
4: Comma (no space after 
comma) 
 
5: Parameter 2 
Parameter 
example 
Boolean 
Boolean logic: 0 or 1. Used for On (1) or 
Off (0) command. 
NR1 
Integer: 0, 1, 2, 3..... 
 NR2 
Decimal number: 0.0, 0.1, 0.2,.... 
 NR3 
Floating point number: 4.5e-1, 8.5e+1,... 
 NRf 
alue available. 
 DEF 
Default setting value. 
Automatic 
parameter range 
selection 
The GDM-8261 automatically translates the command 
parameter into the closest available value. 
Example 1 
CONF:VOLT:DC 1 (Sets the 
measurement item to DC Voltage and the 
range to 1V). 
 Example 2 
CONF:VOLT:DC 2 (Sets the 
measurement item to DC Voltage and the 
range to 2V). There is no 2V range so the 
GDM-8261 selects the closest range, 10V.
EOL 
Marks the end of a command line. The following 
messages are in accordance with IEEE488.2 standard. 
 LF, CR,CR+LF 
EOL, user configurable, see page 128. 
Message 
Separator 
EOL or ,   
Command separator, user configurable 
(excluding GPIB), see page 129.   