GDM-8261 User Manual
88
4. Press the Enter key to confirm
your selection. Press the Exit
key to go back to the default
display.
Gate Time Setting
Background
The gate time settings determine the accuracy of the
frequency and period measurements. The gate time
settings are the equivalent to the Fast, Medium and Slow
rate settings.
Setting
10ms, 100ms, 1000ms
Rate Settings
The gate time settings are analogous to the rate settings.
Function Digits Rate Readings/s Gate time
Frequency,
Period
6 ½ Slow 1 1000ms
5 ½ Med. 10 100ms
4 ½ Fast 100 10ms
Panel operation
1. Press the Shift key followed by
the 2nd (Menu) key. Press the
right key three times. The Hz/P
menu appears.
2ND
MENU
2. Press the Down key, the right
key and then the down key. The
gate time settings menu appears.
1st display
Shows the gate time setting
3. Select the gate time using the
Up/Down key.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com