ADVANCED MEASUREMENT
53
Advanced Measurement Overview
Background
A
dvanced measurement mainly refers to the type of
measurement which uses the result obtained by one of
the basic measurements: ACV, DCV, ACI, DCI, 2/4W,
Diode/Continuity, Frequency/Period, and Temperature.
Advanced
Measurement
Basic Measurement
AC/DCV
AC/DCI 2/4W Hz/P TEMP
dB
— — — — —
dBm
— — — — —
Max/Min
Relative
Hold
Compare
Math
Refresh Rate
Background
Refresh rate defines how frequently the GDM-8261
captures and updates measurement data. A faster refresh
rate yields a lower accuracy and resolution. A slower
refresh rate yields a higher accuracy and resolution.
Consider these tradeoffs when selecting the refresh rate.
For DC measurements, the frequency of the refresh rate
depends on the rate settings (S, M, F) and the ADC speed
settings (Accurate, Quick) (page 86 ).
For AC measurements, the refresh rate (S, M, F) is
directly tied to the AC bandwidth settings (page 80).
For further details, please see the specifications.