190
Temperature Characteristics
(Display in °C, °F, Exclusive of probe errors.)
RTD (Accuracy based on PT100):
(100 platinum [PT100], D100, F100,PT385, PT3916, or user type)
[1]
Range Resolution 1 Year (23°C ±5°C)*
Temperature Coefficient
0°-18°C & 28°-55°C
-200°C~ -100°C 0.001°C 0.09°C 0.004 °C / °C
-100°C~-20°C 0.001°C 0.08°C 0.005 °C / °C
-20°C~20°C 0.001°C 0.06°C 0.005 °C / °C
20°C~100 °C 0.001°C 0.08°C 0.005 °C / °C
100°C~300 °C 0.001°C 0.12°C 0.007 °C / °C
300°C~600 °C 0.001°C 0.22°C 0.009 °C / °C
Thermocouples (Accuracy based on ITS-90)
[1]
:
Type Range Resolution
90 Day/1 Year
(23°C±5°C)*
Temperature Coefficient
0°-18°C & 28°-55°C
E -200 to +1000°C 0.002 °C 0.2 °C 0.03 °C / °C
J
-210 to +1200˚
C
0.002 °C 0.2 °C 0.03 °C / °C
T -200 to +400°C 0.002 °C 0.3 °C 0.04 °C / °C
K
-200 to +1372˚
C
0.002 °C 0.3 °C 0.04 °C / °C
N
-200 to +1300˚
C
0.003 °C 0.4 °C 0.05 °C / °C
R
-50 to +1768˚
C
0.01 °C 1 °C 0.14 °C / °C
S
-50 to +1768˚
C
0.01 °C 1 °C 0.14 °C / °C
B
+350 to +1820˚
C
0.01 °C 1 °C 0.14 °C / °C
[ 1 ] Specifications do not include probe accuracy
*Relative to simulated junction