164
CALCulate:DBM:REFerence
Sets the reference value for the dBm function.
Parameter: <NRf> | MIN | MAX
Example: CALC:DBM:REF MAX
Sets the reference value for dBm measurements to the maximum allowed.
CALCulate:DBM:REFerence?
Returns the reference value from the dBm function.
CALCulate:STORe:COUNt
Set the number of measurement counts that are recorded with the Store
measurement function.
Parameter: <NR1> (2 ~ 9999) | MIN | MAX
Example: CALC:STOR:COUN 1000
Sets the number of counts to be recorded as 1000.
CALCulate:STORe:COUNt?
Returns the number of counts that are recorded with the Store
measurement function.
Parameter: [None] | MIN | MAX
CALCulate:AVERage:COUNt?
Returns the total number of recorded counts.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats
Example: CALC:AVER:COUN? 0
>+0010
Returns the total number of counts set for the Store function (10 counts).
CALCulate:AVERage:MINimum?
Returns the minimum recorded value.
Parameter: None | <NR1>(0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:MAXimum?
Returns the maximum recorded value.
Parameter: None | <NR1>(0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:AVERage?
Returns the average recorded value.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:PTPeak?
Returns the recorded peak to peak value (max value – min value).
Parameter: None | <NR1> (0|1|2) 0=Store, 1=Scan, 2=Stats
Return Parameter: <NRf>
CALCulate:AVERage:SDEViation?
Returns the recorded Standard Deviation.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats