Smartdop
®
30EX Basic Operation
1. INDICATIONS FOR USE, CONTRAINDICATIONS, &
APPLICATIONS
INDICATIONS FOR USE
The Smartdop® 30EX intended for use in the noninvasive evaluation of peripheral vascular
pathology in patients by detecting arterial and venous blood ow in extremities.
CONTRAINDICATIONS
Not intended for use in ophthalmic or periorbital studies as the acoustic intensity is higher than
the level recommended for use on and around the eye.
APPLICATIONS
(Refer to section 2 for Noninvasive Arterial & Venous testing instructions)
1-1 DOPPLER ARTERIAL & VENOUS APPLICATIONS
Ankle Brachial Index (ABI) studies • Systolic pressures
Peripheral vascular procedures Blood ow velocity
Segmental Pressure studies • Penile & digit systolic pressures
Venous compressions
1-2 PPG (PHOTOPLETHYSMOGRAPHY) AND PULSE VOLUME (PV) ARTERIAL &
VENOUS APPLICATIONS
(PPG applications require the optional PG-21 Photoplethysmography probe, supplied
separately)
PPG toe pressures & Toe Brachial Index (TBI)
PPG venous reux
Pulse Volume Recordings (PVR)
2
NOTE
The Smartdop
®
30EX has built-in
pulse volume capabilities. No
additional PV module is required
to perform pulse volume arterial
studies.
NOTE
The Smartdop
®
30EX and associated Doppler probes, cus, and photoplethysmography
(PPG) probe are not intended to be placed in contact with blood or mucous membranes.
NOTE
The Smartdop
®
30EX is not recommended for fetal use.