Trouble shooting 147
ORIG0140-147
4127718F/B
02/26/2007
SOLUTIONPROBLEM
R152 - RTD overheating
printhead
Call services
R160 - Communication error
Check Cables
R170 – Invalid scale for RTD.
Choose another scale
R180 – Signature error.
Call services
Step 1: check if the RTD is close to
a heat source or is exposed to di-
rect sunlight.
Step 2: if yes, move the RTD.
Step 3: wait 15 mn to cool down.
Step 1: turn Power machine off and
on . If after 2 attempts the fault is not
cleared then,
Step 2: check if the cable is correctly
connected; if not,
Step 3: change the cable or ask for
RTD exchange procedure.
Step 1: switch OFF/ON the RTD to
reload scale.
Step 2: retry to print a "0" value
label to confirm.
Step 3: if not, choose another scale.
Turn Power machine off and on. If
after 2 attempts the fault is cleared
then, print a zero value label to
confirm.
If the fault is still not cleared, then
ask for RTD exchange procedure.