Configure channel parameters in HBK ADVANTAGE
Configure channel parameters in HBK ADVANTAGE
Configure a bridge board (B201-10)
Configure each bridge board channel with the following parameters:
l
Channel: The channel name. This is initially derived from the device, but may be renamed.
l
Value: The value of the signal with its engineering unit.
l
Sample rate: The number of samples per second the signal delivers.
l
Low pass: You can eliminate noise with low-pass filters that let frequencies below a specified
threshold pass and eliminate the frequencies above it. Select Off (filter not applied), Butterworth or
Bessel then a value (for example: Butterworth and 20Hz).
NOTE
Ensure that the domain and the filter/sample rate combination is supported for your hardware
configuration. See "Sample rate" on page46 for more information.
If the filter/sample rate combination is not supported a notification message is displayed in the
software and the previous settings are used.
l
Sensor: The sensor type connected to the channel. Only supported sensor types for the connected
device are listed.
l
Zero offset: (in the table only) The value that is subtracted from the measured values before recording
to zero balance the channel.
Configure a CCLD board (B301D-12)
Configure each CCLD board channel with the following parameters:
l
Channel: The channel name. This is initially derived from the device, but may be renamed.
l
Value: (in the table only) The value of the signal in the default engineering unit.
l
Sample rate: The number of samples per second delivered by the signal.
l
Low pass: Not applicable for a CCLD board.
l
High pass: You can eliminate noise with high-pass filters that let frequencies above a specified
threshold pass and eliminate the frequencies below it. Select 0.05, 0.1, 0.7, 1.0, 7.0 or 22.4 Hz.
l
Sensor: Select the sensor type connected to the channel, either CCLD (for CCLD microphones or
accelerometers) or Voltage.
l
Zero offset : (in the table only) Not applicable for a CCLD board.
- 40 -