EasyManua.ls Logo

HBK FUSION - Configure channel parameters in HBK ADVANTAGE; Configure a bridge board (B201-10); Configure a CCLD board (B301 D-12)

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configure channel parameters in HBK ADVANTAGE
Configure channel parameters in HBK ADVANTAGE
Configure a bridge board (B201-10)
Configure each bridge board channel with the following parameters:
l
Channel: The channel name. This is initially derived from the device, but may be renamed.
l
Value: The value of the signal with its engineering unit.
l
Sample rate: The number of samples per second the signal delivers.
l
Low pass: You can eliminate noise with low-pass filters that let frequencies below a specified
threshold pass and eliminate the frequencies above it. Select Off (filter not applied), Butterworth or
Bessel then a value (for example: Butterworth and 20Hz).
NOTE
Ensure that the domain and the filter/sample rate combination is supported for your hardware
configuration. See "Sample rate" on page46 for more information.
If the filter/sample rate combination is not supported a notification message is displayed in the
software and the previous settings are used.
l
Sensor: The sensor type connected to the channel. Only supported sensor types for the connected
device are listed.
l
Zero offset: (in the table only) The value that is subtracted from the measured values before recording
to zero balance the channel.
Configure a CCLD board (B301D-12)
Configure each CCLD board channel with the following parameters:
l
Channel: The channel name. This is initially derived from the device, but may be renamed.
l
Value: (in the table only) The value of the signal in the default engineering unit.
l
Sample rate: The number of samples per second delivered by the signal.
l
Low pass: Not applicable for a CCLD board.
l
High pass: You can eliminate noise with high-pass filters that let frequencies above a specified
threshold pass and eliminate the frequencies below it. Select 0.05, 0.1, 0.7, 1.0, 7.0 or 22.4 Hz.
l
Sensor: Select the sensor type connected to the channel, either CCLD (for CCLD microphones or
accelerometers) or Voltage.
l
Zero offset : (in the table only) Not applicable for a CCLD board.
- 40 -