Averaging Feature
Details of Operation
Calculating a moving average of displayed resistance values
Initial Status
OFF
Configuration Method
Change number of times averaging is conducted by pressing the AVG
key
OFF (once) 4 times 8 times 16 times OFF …
Zero-Adjustment
Details of Operation
Measured values upon implementation (upon obtaining adjusted
values) are set at zero
Initial Status
OFF
Adjustment Range
Up to 300 counts for each range (resistance, voltage)
Auto Holding of Measured Values
Details of Operation
Holding (ceasing updating of) displayed values
Hold Method
1. Pressing the HOLD key
2. Inputting signals to the EXT.HOL/MEMO terminal
3. Stabilising measured values (when the auto-hold feature is on)
Comparator Feature
Details of Operation
Comparison of measured values with permissible values
Initial Status
OFF
Setting Method
Press the COMP key to select the comparator no. to use.
Configure resistance upper limit no.1, resistance upper limit no.2, and
the voltage lower limit.
Resistance
(Low)
Resistance
(Medium)
Resistance
(High)
Voltage (High)
PASS
WARN
FAIL
Voltage (Low)
WARN
WARN
FAIL
Settings Saved