121
BasicSpecications
9.2 Basic Specications
Measurement
parameters
• Measuring battery internal resistance
• Measuring battery terminal voltage (DC voltage only)
• Measuring temperature
Measurable
ranges
Resistance measurement:
0.000 m
Ω
to 3.100
Ω
(4-rangeconguration)
Voltage measurement:
0.000Vto±60.00V(2-rangeconguration)
Temperature measurement:
−10.0°Cto60.0°C(single-rangeconguration)
Maximum input
voltage
60 V DC (between the positive and negative
measurement terminals)
AC voltage cannot input.
Maximum rated
voltage to earth
60 V DC (Measurement category not rated)
Anticipated transient overvoltage 330 V (between all
measurement terminals and ground)
Measurement
method
Resistance measurement:
AC four-terminal method
Open-circuited terminal voltage:
5 V peak at a maximum
Measurement current:
1.6 mA to 160 mA (Fixed according to the
resistance measurement range setting)
Temperature measurement:
Platinum temperature sensor (500
Ω
at 25°C)
A/D conversion method:
Delta-sigma modulation
Display update rate:
3 times/s (Resistance, voltage, and temperature
measured as a set)
HIOKI BT3554F961-00HIOKI BT3554F961-00