65
Measuring Batteries (Inspection)
Measurement error
Even if the [−−−−] segments appear and the [OVER] segment
blinks on the display (at the same time, the maximum display value
blinks), this does not indicate an error.
[−−−−] If the resistance readout exhibits the [----] segments,
the test lead may be open-circuited.
Otherwise, the instrument cannot perform
measurement because of a failure, such as no current
ows due to a broken test lead.
The test lead may not be connected correctly to the
target under measurement.
The resistance of the target under measurement
signicantly exceeds the measurement range.
Blinking the
[OVER] segment
and maximum
display value
Resistance, voltage, or temperature may exceed each
measurable range.
IMPORTANT
When measuring contact resistances of relays or connectors, be
aware that the instrument will generate an open-terminal voltage
of 5 V at maximum. The open-terminal voltage may damage the
oxidized coating on the contacts of targets under measurement,
leading to incorrect measurements.
Warning display
If an overvoltage is input, the instrument will show the [OVER]
segment and the blinking maximum display value on the display,
backlight the display in red, and produce beeps.