AccuracySpecications
9.3 Accuracy Specications
Accuracy
guarantee
conditions
Accuracy guarantee duration:
1 year
Accuracy guarantee duration after adjustment made by
Hioki:
1 year
Accuracy guarantee temperature and humidity range:
23°C ±5°C (73°F ±9°F), 80% RH or less
Warm-up time:
None (Not required)
Temperature
characteristics
Within the operating temperatures, add (measurement
accuracy) × 0.1/°C.
(Except the range of 18°C to 28°C)
Resistance
measurement
accuracy
Measurement current accuracy: ±10%
Measurement current frequency: 1 kHz ±30 Hz
1 kHz ±80 Hz if the noise-frequency reduction function
is enabled.
Range
setting
Maximum
displayable
value
Resolution Measurement accuracy
Measurement
current
3 m
Ω
3.100 m
Ω
1 µ
Ω
±1.0% of reading
±8 digits
160 mA
30 m
Ω
31.00 m
Ω
10 µ
Ω
±0.8% of reading
±6 digits
160 mA
300 m
Ω
310.0 m
Ω
100 µ
Ω
16 mA
3
Ω
3.100
Ω
1 m
Ω
1.6 mA
• Add the following values if zero adjustment has not been performed.
When the 9465-10 is used: ±5 digits
When L2020 is used: ±6 digits
When the 9772 is used: ±1 digit
When the 9460 is used: ±16 digits
When the 9467 is used: ±5 digits
HIOKI BT3554F961-00HIOKI BT3554F961-00