Adjusting the Zero Point (Zero Adjustment)
Performing zero adjustment
1
Check if the test lead has been
connected properly.
Disconnect the test lead from the
measurement target if connected.
(Black)
1
2
(Red)
2
Press the 0ADJ key.
The instrument will be put
on standby for acquiring the
correction value.
3
While the [0AdJ] segment is
blinking, short-circuit the test
lead using the zero-adjustment
board.
See “How to short-circuit various
test lead” (p. 51).
Failure to short-circuit the test
lead while the display is blinking
will result in an error.
Blinks for
about 10 s.
For Pin Type Lead