163
Calibration
Calibrating the voltage measurement unit
Use a generator that can output a DC voltage of 60 V.
Connect the instrument and generator as shown below:
SOURCE
SENSE
SENSE
SOURCE
BT3554-50
DC voltage generator
+
+
+
Do not apply AC current from the instrument to the generator.
Doing so may cause the generator to malfunction.
Use a generator having a low output-impedance (50
Ω
or less).
If the [−−−−] segments are displayed, the wire-break detection
function of the instrument needs to be disabled.
HIOKI BT3554F961-00HIOKI BT3554F961-00