164
Calibration
How to disable the wire-break detection function
1
Turn o󰀨 the instrument.
2
While holding down the A.HOLD/A.MEM key, turn on the
instrument.
The [on] segments will blink.
3
Use the cursor keys to switch over form the [on] segments
to the [oFF] segments.
4
Press the ENTER key.
The wire-break detection function will be disabled and the
instrument will be restarted.
Restart the instrument after calibration. The wire-break detection
function will be enabled again. Do not disable the wire-break
detection function in normal use for measurement.
HIOKI BT3554F961-00HIOKI BT3554F961-00