Operating Precautions
Precautions for measurement
DANGER
To prevent an electric shock, be careful to avoid
shorting live lines with the test lead tip.
WARNING
Do not use the instrument to measure circuits that
exceed its ratings or specications. Damage to the
instrument can cause an electric shock.
Do not measure any voltage that would exceed the
instrument’s maximum input voltage (terminal-to-
terminal) or maximum rated line-to-ground voltage
of 60 V.
Maximum input voltage
(terminal-to-terminal)
60 V DC
Maximum rated
voltage to earth
60 V DC
Do not measure AC voltage.