Convenient functions
Comparator function
You can set threshold values to evaluate battery deterioration using
the comparator function. (p. 67)
Example of setting deterioration evaluation values
Acceptance value
Resistance: 0.5 m
Ω
Voltage: 2.0 V
Warning threshold
0.75 m
Ω
(Fresh battery’s value × 1.5)
1.8 V
(Fresh battery’s value × 0.9)
Rejection value
1.0 m
Ω
(Fresh battery’s value × 2)
PASS WARNING FAIL
Fresh In use
Need
replacing
Hold down for
at least 1 s.
Downloading measured values to your computer
You can connect the instrument and your computer using the
accompanying USB cable to download measured values. (p. 103)