Adjusting the Zero Point (Zero Adjustment)
Keep the zero-adjustment board at least 10 cm away from the
instrument.
Always use the accompanying or optional zero-adjustment
board when performing zero adjustment.
Insert the pin tips into the hole on the zero-adjustment board
and place each of the SOURCE and SENSE terminals into
contact with the hole on the board. (See the gure below.)
SOURCE
SENSE
Do not place the zero-adjustment board on top of the battery
or any metal. Electromagnetic induction e󰀨ect could result in
unstable measurement values. In such a case, keep the zero-
adjustment board away from any metal.
If zero adjustment is performed with the Pin Type Lead short-
circuited at the tips or using a metal sheet other than the zero-
adjustment board intended for this purpose, the instrument
cannot adjust the zero point accurately.
When the distance between the terminals on the battery (target
under measurement) is more than that between the holes on
the zero-adjustment board, use the outermost holes to perform
zero adjustment.
Consider the zero-adjustment board to be a consumable. It
is recommended to replace it with a new one after the use of
around 700 times.