73
Setting Threshold Values for the Comparator
7
Set the voltage evaluation
method.
When you have chosen the
[PoL], reversely connecting the
red and black wires of the test
lead will cause the [WARNING]
result.
(Black)(Red) (Red)(Black)
[WARNING]
Voltage
evaluation
method
Comparison method
Data to be
saved
[AbS]
(Default
setting)
Evaluates absolute voltage
values, regardless of whether
they are positive or negative.
Signed
(minus sign
only)
[PoL]
The [WARNING] result will be
given for the negative voltage
value.
If the test lead is placed into
contact in reverse polarity with
a battery (the red and black
probes with the negative and
positive terminals, respectively),
the [WARNING] result will be
given.
Signed
(minus sign
only)
You can set the evaluation method using version 1.8 or
later of GENNECT Cross.
HIOKI BT3554F961-00HIOKI BT3554F961-00