130
6.2 Sensor Data
6.2.1 Operating and Storage Conditions for Performance Tested EC Cartridges
Gas
Cartridge
P/N
Operating Pres-
sure
Operating Air
Speed
Warm-up Time
(minimum)
Storage Conditions*
Temperature Pressure Humidity Time**
O
2
Oxygen
XNXXSO1SS
80kPa~120kPa 0~6m/sec 60sec. 0to20°C,32to68°F 80to120KPa 5to95%RH 6months
XNXXSO1FM
H
2
S HydrogenSulfide
XNXXSH1SS
80kPa~120kPa 0~6m/sec 60sec. 0to20°C,32to68°F 70to110KPa 30to70%RH 6months
XNXXSH1FM
H
2
S(High) HydrogenSulfide XNXXSH2SS 80kPa~120kPa 0~6m/sec 60sec. 0to20°C,32to68°F 70to110KPa 30to70%RH 6months
CO CarbonMonoxide
XNXXSC1SS
80kPa~120kPa 0~6m/sec 60sec. 0to20°C,32to68°F 70to110KPa 30to70%RH 6months
XNXXSC1FM
*Storeinsealedpackages**Checkcartridgecertificates

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
Ingress ProtectionIP66/67
Product TypeUniversal Transmitter
Enclosure RatingIP66/67
DisplayLCD
Communication ProtocolsHART, Modbus, Foundation Fieldbus
Power Supply24 VDC nominal, 18-32 VDC
Operating Temperature-40°C to +65°C (-40°F to +149°F)
MaterialAluminum or Stainless Steel
Gas TypesToxic, Oxygen, and Combustible gases
Output4-20 mA
Relays3 programmable relays
ApprovalsATEX, IECEx, UL, CSA
Sensor CompatibilityElectrochemical, catalytic, infrared, and other sensor types

Related product manuals