EasyManua.ls Logo

Honeywell XNX - Zero Calibration

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XNX Universal Transmitter
Section 3 - Calibration
91
3.2.1 Zero and Span Calibration for XNX EC Sensors,
mV Sensors, and Searchpoint Optima
Caution: Beforeinitialcalibration,allowthesensortostabilizefor30minutes
afterapplyingpower.WheninZeroandSpanCalibrationmodes,thecurrent
outputfromthesensorisinhibited(default2mA)toavoidfalsealarms.
Caution:Formoststickygases(e.g.,HCl,Cl
2
)usePTFEtubingwithshort
piecesofrubbertubeforthefinalconnection(duetotheinflexibilityofPTFE).This
minimizesadhesionofthegastothetubesurfaceandallowsmoreaccurate
measurement.
Tocalibratethesensor,useanappropriatespangascylinder,
tubing,magnet,andcalibrationgasowhousing�Settheow
regulatorto300-375ml/minforXNXECsensorsor500±200ml/
minforXNXmVsensors�Acompressedgascylinder(20�9%Vol
oxygen)shouldbeusedtoperformthezerocalibrationifthe
areawherethesensorislocatedcontainsanyresidualamount
ofthetargetgas�Ifnoresidualgasispresent,backgroundair
canbeusedtoperformthezerocalibration�ContactaHoneywell
Analyticsrepresentativefordetailsaboutsuitablecalibrationkits�
Tocalibratethesensor,followtheprocedureinSection3�2�2
Note: Theoxygensensordoesnotrequireazeroingprocedure.Background
air(20.9%Voloxygen)canbeusedtospantheoxygensensorinplaceofa
compressedaircylinder(20.9%Voloxygen).SeeSection6.3.2forothersensors.
Note: ENperformancestandardsrequire10minutesstabilizationtimefor
applicationofzeroandspangasforperformance-approvedEC,mV,andIR
sensorspriortocalibration.
3.2.2 Calibration Procedure
Thissectionoutlinesthestepsforcalibratingthetransmitter’s
attachedsensors�
Note: TheZeroCalibrationprocedureshouldbeperformedpriortotheSpan
Calibrationprocedure.
1�Ifusingacompressedgascylinder,pushthecalibration
gasowhousingontothebottomofthesensorandapply
thegas�
2�AccesstheGasCalibrationMenu�
Figure 194. Gas Calibration Menu
Note: TheGasCalibrationmenuisforbothZeroCalibrationandSpan
Calibration.
Zero Calibration
Sensor Reading
at Current Settings
Figure 195. Zero Calibration Screen

Table of Contents

Other manuals for Honeywell XNX

Related product manuals