EasyManuals Logo

Honeywell XNX Technical Manual

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
XNX Universal Transmitter
Section 3 - Calibration
91
3.2.1 Zero and Span Calibration for XNX EC Sensors,
mV Sensors, and Searchpoint Optima
Caution: Beforeinitialcalibration,allowthesensortostabilizefor30minutes
afterapplyingpower.WheninZeroandSpanCalibrationmodes,thecurrent
outputfromthesensorisinhibited(default2mA)toavoidfalsealarms.
Caution:Formoststickygases(e.g.,HCl,Cl
2
)usePTFEtubingwithshort
piecesofrubbertubeforthefinalconnection(duetotheinflexibilityofPTFE).This
minimizesadhesionofthegastothetubesurfaceandallowsmoreaccurate
measurement.
Tocalibratethesensor,useanappropriatespangascylinder,
tubing,magnet,andcalibrationgasowhousing�Settheow
regulatorto300-375ml/minforXNXECsensorsor500±200ml/
minforXNXmVsensors�Acompressedgascylinder(20�9%Vol
oxygen)shouldbeusedtoperformthezerocalibrationifthe
areawherethesensorislocatedcontainsanyresidualamount
ofthetargetgas�Ifnoresidualgasispresent,backgroundair
canbeusedtoperformthezerocalibration�ContactaHoneywell
Analyticsrepresentativefordetailsaboutsuitablecalibrationkits�
Tocalibratethesensor,followtheprocedureinSection3�2�2
Note: Theoxygensensordoesnotrequireazeroingprocedure.Background
air(20.9%Voloxygen)canbeusedtospantheoxygensensorinplaceofa
compressedaircylinder(20.9%Voloxygen).SeeSection6.3.2forothersensors.
Note: ENperformancestandardsrequire10minutesstabilizationtimefor
applicationofzeroandspangasforperformance-approvedEC,mV,andIR
sensorspriortocalibration.
3.2.2 Calibration Procedure
Thissectionoutlinesthestepsforcalibratingthetransmitter’s
attachedsensors�
Note: TheZeroCalibrationprocedureshouldbeperformedpriortotheSpan
Calibrationprocedure.
1�Ifusingacompressedgascylinder,pushthecalibration
gasowhousingontothebottomofthesensorandapply
thegas�
2�AccesstheGasCalibrationMenu�
Figure 194. Gas Calibration Menu
Note: TheGasCalibrationmenuisforbothZeroCalibrationandSpan
Calibration.
Zero Calibration
Sensor Reading
at Current Settings
Figure 195. Zero Calibration Screen

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
Ingress ProtectionIP66/67
Product TypeUniversal Transmitter
Enclosure RatingIP66/67
DisplayLCD
Communication ProtocolsHART, Modbus, Foundation Fieldbus
Power Supply24 VDC nominal, 18-32 VDC
Operating Temperature-40°C to +65°C (-40°F to +149°F)
MaterialAluminum or Stainless Steel
Gas TypesToxic, Oxygen, and Combustible gases
Output4-20 mA
Relays3 programmable relays
ApprovalsATEX, IECEx, UL, CSA
Sensor CompatibilityElectrochemical, catalytic, infrared, and other sensor types

Related product manuals