EasyManua.ls Logo

Honeywell XNX - Calibration Interval

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XNX Universal Transmitter
Section 2 - Installation and Operation
71
Searchline(seethetableinthenextcolumn)�
Figure 102. mA Levels Menu
Usingthe switches,movetothemAoutputtobechanged
anduse
toselectit�
Figure 103. Set mA Levels for Warning
Usethe switchestodecrementorincrementthevalueuntil
thedesiredvalueappears�Use
toselectthevalueandmove
tothenextsetting�Repeatforeachsettingtobechanged�
ThedefaultvaluesandavailableoutputrangesforInhibit,
Warning,Overrange,BeamBlocked,andLowSignalareshown
inthefollowingtable�SeeSection5Warnings/Faultsformore
information�

Signal
Output (mA)
Default Min Max
I
Inhibit 2.0 1.0 3.5
W
Warning 3.0 1.0 3.5
O
Overrange 21.0 20 22
B
BeamBlocked 1.0 1.0 4.0
L
LowSignal 1.0 1.0 4.0
Figure 104. Set mA Levels for Inhibit
Afterallchangeshavebeenmade,usethe switchesto
movetothe‘ü’anduse
onthefrontpaneltoacceptandsave
thesettings�If ü’isnotselected,noneofthechangeswillbe
saved�
Figure 105. mA Settings Saved
Calibration Interval
CalibrationIntervalallowsadesiredintervalforsensor
calibrationtobesetforsensorsattachedtothetransmitter�The
transmitterwillgenerateawarningwhentheintervalisreached�
Figure 106. Calibration Interval Menu
CalibrationIntervalwillnotappearwhenanIRpersonalityboard
isattachedandthemAsensortypeissetas‘OthermASensor’�
Thedefaultcalibrationvaluesforthe“CalibrationRequired”
diagnosticvarybasedonsensortype�Thisvaluecanberepro-
grammedinaccordancewithsiterequirementstoensurethe
highestlevelofsafety�Correctoperationofeachsensorshould
beconrmedusingcalibrationwithacertiedgasofknowncon-
centrationbeforecommissioning�
AlthoughthecalibrationIntervalcanbesettoanyvaluebe-
tween0and360days,HoneywellAnalyticsrecommendsthatthe
intervalforelectrochemicalandcatalyticsensorsbesetto180

Table of Contents

Other manuals for Honeywell XNX

Related product manuals