EasyManuals Logo

Honeywell XNX Technical Manual

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
XNX Universal Transmitter
Section 2 - Installation and Operation
71
Searchline(seethetableinthenextcolumn)�
Figure 102. mA Levels Menu
Usingthe switches,movetothemAoutputtobechanged
anduse
toselectit�
Figure 103. Set mA Levels for Warning
Usethe switchestodecrementorincrementthevalueuntil
thedesiredvalueappears�Use
toselectthevalueandmove
tothenextsetting�Repeatforeachsettingtobechanged�
ThedefaultvaluesandavailableoutputrangesforInhibit,
Warning,Overrange,BeamBlocked,andLowSignalareshown
inthefollowingtable�SeeSection5Warnings/Faultsformore
information�

Signal
Output (mA)
Default Min Max
I
Inhibit 2.0 1.0 3.5
W
Warning 3.0 1.0 3.5
O
Overrange 21.0 20 22
B
BeamBlocked 1.0 1.0 4.0
L
LowSignal 1.0 1.0 4.0
Figure 104. Set mA Levels for Inhibit
Afterallchangeshavebeenmade,usethe switchesto
movetothe‘ü’anduse
onthefrontpaneltoacceptandsave
thesettings�If ü’isnotselected,noneofthechangeswillbe
saved�
Figure 105. mA Settings Saved
Calibration Interval
CalibrationIntervalallowsadesiredintervalforsensor
calibrationtobesetforsensorsattachedtothetransmitter�The
transmitterwillgenerateawarningwhentheintervalisreached�
Figure 106. Calibration Interval Menu
CalibrationIntervalwillnotappearwhenanIRpersonalityboard
isattachedandthemAsensortypeissetas‘OthermASensor’�
Thedefaultcalibrationvaluesforthe“CalibrationRequired”
diagnosticvarybasedonsensortype�Thisvaluecanberepro-
grammedinaccordancewithsiterequirementstoensurethe
highestlevelofsafety�Correctoperationofeachsensorshould
beconrmedusingcalibrationwithacertiedgasofknowncon-
centrationbeforecommissioning�
AlthoughthecalibrationIntervalcanbesettoanyvaluebe-
tween0and360days,HoneywellAnalyticsrecommendsthatthe
intervalforelectrochemicalandcatalyticsensorsbesetto180

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
Ingress ProtectionIP66/67
Product TypeUniversal Transmitter
Enclosure RatingIP66/67
DisplayLCD
Communication ProtocolsHART, Modbus, Foundation Fieldbus
Power Supply24 VDC nominal, 18-32 VDC
Operating Temperature-40°C to +65°C (-40°F to +149°F)
MaterialAluminum or Stainless Steel
Gas TypesToxic, Oxygen, and Combustible gases
Output4-20 mA
Relays3 programmable relays
ApprovalsATEX, IECEx, UL, CSA
Sensor CompatibilityElectrochemical, catalytic, infrared, and other sensor types

Related product manuals