EasyManua.ls Logo

Honeywell XNX - 3�2�3 Using The Calibration Cup

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XNX Universal Transmitter
Section 3 - Calibration
93
Figure 200. Span Passed Screen
Ifthecalibrationisnotsuccessful,theSpanFailedscreen
willdisplay�Selecting
willreturntotheSpanConcentration
screentobeginthespancalibrationagain�Selecting
will
exitSpanCalibrationandreturntotheGasCalibrationMenu�
Arrow indicates
gas values too low
Selecting ‘ returns to
the Span Concentration
screen
Figure 201. Span Calibration Failed
OncetheZeroGasandSpancalibrationsarecompleted
successfully,theuserwillbepromptedto:
•exitwithinhibitoff
•exitwithinhibiton,or
•notexit�
Figure 202. Span Calibration Failed
Warning: WhentheXNXtransmitterisplacedinInhibitMode,alarmsare
silenced.Thiswillpreventanactualgaseventfrombeingreported.InhibitMode
mustbelimitedtotestingandmaintenanceonly.ExitInhibitModeaftertesting
ormaintenanceactivities.
3.2.3 Using the Calibration Cup
RefertoFigure203toattachthecalibrationcup:
1� Snapthecalibrationcupintotheweatherprotector�The
twoprotrusionsonthecuptintorecessesintheweather
protector
2� Attachthehosefromthegascylindertothecalibrationcup�
Notethatthecup’sowisunidirectional�Thereisanarrow
onthebottomshowingowdirection
3� Adjustthecalibrationowrate
1
weather
protector
calibration cup
attach gas
cylinder hose
Figure 203. Attaching the Calibration Cup
1300-375ml/minforXNXECsensors,500±200ml/minforXNXmVsensors,unless
otherwisedirected

Table of Contents

Other manuals for Honeywell XNX

Related product manuals