EasyManua.ls Logo

Honeywell XNX - 3�2�5 705;705 Ht Calibrating; 3�2�7 Calibrating The Searchpoint Optima Plus; Sulfide (H S) Sensors

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XNX Universal Transmitter
Section 3 - Calibration
94
3.2.4
Zero and Span Calibration of XNX EC Hydrogen
Sulde(H
2
S) Sensors
Caution: Beforeinitialcalibration,allowthesensortostabilizefor30minutes
afterapplyingpower.Wheninzeroandspancalibrationmodes,thecurrent
outputfromthesensorisinhibited(default2mA)toavoidfalsealarms.
Caution: Recalibrateifthetemperatureoflocalenvironmenthasvariedby
morethan±15°Cfromthetemperatureofcalibration.
HydrogenSuldesensorscanbeaffectedbyextremehumidity
changes�Asuddenincreaseinambienthumiditycanresultina
short-termpositivedriftintheinstrument’sreading�Asudden
decreaseinambienthumiditycanresultinashort-termnegative
driftintheinstrument’sreading�Thesearemostlikelytobe
noticedduringcalibrationwithdryorcylindergas�
Whencalibratinghydrogensuldecartridges,thefollowing
shouldbetakenintoaccountwhilefollowingtheprocedurein
Section3�2�2�
1�Tozerothesensor,useacompressedgascylinderof
20�9%Voloxygen(notnitrogen)�Donotusebackground
air
Ifaspancalibrationistobeperformed,thespancalibrationgas
shouldbeappliedtothesensorimmediatelyafterthezeroing
procedure�Donotallowthesensortoreturntoambientair
conditions�
Warning: Long-termexposure(>20minutes)toconcentrationsexceediingthe
full-scalerangeofthesensorcancauseittolosesensitivity.Theouputofthe
sensormaythendecreaseinvalueeventhoughhighlevelsoftoxicgasarestill
present.Beforere-calibratingthetransmitter,verifytheabsenceofgas.
3.2.5 705/705HT Calibrating
Forcompletecalibrationandcongurationinformation,seethe
Type705OperatingInstructions(P/N:00705M5002)�
3.2.6 Sensepoint/Sensepoint HT Calibrating
Forcompletecalibrationandcongurationinformation,seethe
SiegerSensepointTechnicalHandbook(P/N:2106M0502)�
3.2.7 Calibrating the Searchpoint Optima Plus
Completecalibrationandcongurationinformationcanbe
foundintheSearchpointOptimaPlusOperatingInstructions
(P/N:2108M0501)�Ifproperlyinstalledandmaintained,the
SearchpointOptimaPlussensorwillnotrequireroutine
calibration�ThisisduetotheinherentstabilityoftheIR
absorptionprocessandtheunit’sfullycompensatedoptical
conguration�
1� FromtheCalibrationmenu,selecttheGasCalibrationoption�
Figure 204. Calibration menu

Table of Contents

Other manuals for Honeywell XNX

Related product manuals