SCPI
COMMAND SUMMARY
a
*OPT?
This returns a string identifying any device options.
0
*RCL
<num>
The instrument state is recalled from the specified memory register.
The value range is from 0 to 8.
l
*RST
The synthesizer is set to a predefined condition as follows:
AM:DEPTH value is 50%
AM INTernal FREquency value is 1
kHz
AM:MODE
NORMal
AM:SOURce
EXTernal
AM:STATe OFF
AM:TYPE
LINear
CALibration:PEAKing:AUTO OFF
CALibration:POWer:ATTenationO DBM
CALibration:POWer:RANGel
CALibration:SPAN:AUTO OFF
CORRection:FLATness? returns
a0
CORRection:ARRay
clear
CORRection:FLATness:POINts? returns
a0
CORRection:STATe OFF
DIAGnostics:ABUS:AVERage 1
DIAGnostics:TEST:ENABle ALL
DIAGnostics:TEST:LOG:SOURceFAIL
DIAGnostics:TEST:LOG[:STATel
OFF
DIAGnostics:TEST:LOOP OFF
DISPlay[:STATel
ON
FM:DEViationvalueis
1 MHz
FM:COUPlingAC
FM:FILTer:HPASs MAXimum
FM:INTernal:FREquency
valueis
1 MHz
FM:SENSitivity MAXimum
FM:SOURce EXTernal
FM:STATe OFF
FREquency:CENTer value is
(MAX+MIN)/2
FREQuency:CW value is
(MAX+MIN)/2
FREQuency:CW:AUTO
OFF
FREQuency:MANualvalueis
(MAX+MIN)/2
FREQuency:MODECW
FREQuency:MULTiplier
1
FREQuency:MULTiplier:STATe
OFF
FREQuency:OFFSet
0
FREQuency:OFFSet:STATe
OFF
FREQuency:STARtMINimum
FREquency:STEP
calculated from span
FREQuency:STEP:AUTO
ON
FREQuency:STOP
MAXimum
HP 8360
User’s Handbook
Operating and Programming Reference
S-15