47
2.  Use the down arrow button to move to Select Flow 
Sensor, and press the + button to choose it.
3.  Use the + button to advance through the pipe 
sizes shown in the Fitting Model column until the 
correct size is displayed. The last selection after the 
standard FCT models is OTHER.
4.  If the controller has a decoder output module, and 
the flow meter will be connected to an ICD-SEN 
sensor decoder on the two-wire path, use the down 
arrow to move to Location. Change Location to ADM 
with the + or - buttons.
5.  Exit the dial position to save the settings, unless you 
have selected OTHER.
Additional settings for OTHER
Most other brands of flow sensor require two settings 
for calibration, the K-factor and the Offset. The correct 
settings for these values are found in the sensor 
manufacturer’s documentation, and they are based on the 
pipe type and size.
1.  Consult the “Other” flow sensor’s documentation for 
the correct values for a given pipe size.
2.  If Other is chosen, use the down arrow to move to 
the K-factor setting.
3.  Use the right arrow to skip over any places that need 
to be left at “0”.
4.  At any digit which needs to be changed, use the 
+/- buttons to enter the digit. Then move to the 
next place and repeat until the correct K-factor is 
displayed.
5.  Use the down arrow to move to the Offset value, and 
repeat.
When the correct values are shown for both K-factor and 
Offset, exit the dial position to save the settings. The flow 
sensor will now be calibrated for the pipe size.
SSG (SIMULTANEOUS STATION GROUP) SETUP 
Turn the dial to the Set Program Overlap Options position
See SETTING PROGRAM OVERLAP OPTIONS on page 36 for 
more information. 
SSGs are groups of 2, 3, or 4 stations which are grouped 
together electronically. They will then all run together and 
are programmed as a single unit. Up to 20 SSGs can be 
created in the ACC controller.
This powerful feature can be used to balance flow, shorten 
the water window, simplify programming, and expedite 
common tasks. SSGs can be included in automatic 
programs or custom manual programs, and can be started 
manually from controller at any time. They can also be 
named to make using them easier.
SSGs are not required to operate the ACC controller. They 
are a valuable extra feature for advanced users.
 See SSG Rules on page 49 for more information.