8 - 2
PN 074-519-P1C
3000 Micro GC Operating Manual
Figure 8-1 Example bakeout method
3 Click File >> Method >> Save to save the bakeout method. When prompted,
type a unique name for the bakeout method.
4 Verify that the carrier gas is on and flowing. This will protect the column and
TCD.
5 Click Control >> Download Method to download the bakeout method.
6 Allow the method to run for the duration listed in Table 8-1.
7 After the bakeout is complete, load an analytical method and run a set of
calibration samples. Refer to section 7.7 on page 7-19 and section 7.12 on
page 7-36.
8 Check the report. Adjust the calibration settings, retention times, and method
settings as needed.