10 - 30
IPN 074-505-P1E
IC6 Operating Manual
44
(0x2c)
Ramp 2 Time
(seconds)
(mm:ss)
0/ 5999/ 0
0/ 99:59/ 0
Time
See Time Note
45
(0x2d)
RateWatcher
Option
0/ 1/ 0 Encode 0 = No, 1 = Yes
46
(0x2e)
RateWatcher Time
(seconds)
(mm:ss)
0/ 5999/ 0
0/ 99:59/ 0
Time
See Time Note Grey font unless
RateWatcher option set
to Yes.
47
(0x2f)
RateWatcher
Accuracy (%)
1/ 99/ 5 Integer
48
(0x30)
Multipoint 0/ 1/ 0 Encode 0 = No, 1 = Yes Update not allowed if
material is running.
49
(0x31)
Aggregate Multiplier 0/ 1/ 0 Encode 0 = No, 1 = Yes Visible only if Multipoint
is set to Yes
50
(0x32)
Sensor 1
Failure Action
0/ 4/ 2
Encode
0 = Unused
1 = PostDp
2 = Stop
3 = Suspnd
4 = TimePw
Visible only if Multipoint
is set to Yes
At least one must be
non-zero (when
multipoint set to "Yes")
Update not allowed
when transfer or
backup sensor is
running.
51
(0x33)
Sensor 2
Failure Action
0/ 4/ 0
52
(0x34)
Sensor 3
Failure Action
53
(0x35)
Sensor 4
Failure Action
54
(0x36)
Sensor 5
Failure Action
55
(0x37)
Sensor 6
Failure Action
56
(0x38)
Sensor 7
Failure Action
57
(0x39)
Sensor 8
Failure Action
Table 10-9 Material parameters (continued)
QM
&
UM
Cmnd
ID
Name
(Units)
Range:
Low Limit/
High Limit/
Default
Data Type
Notes
Update Restrictions
and Display Notes