EasyManua.ls Logo

Inficon IC6

Inficon IC6
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 - 47
IPN 074-505-P1E
IC6 Operating Manual
4
(0x04)
Fundamental
Frequency
8 or 64 <Long> or
<Long><Long><Long>
<Long><Long><Long>
<Long><Long>
One long integer for each sensor.
Unconverted. To convert to Hz, multiply by
0.000873114913702011
5
(0x05)
Fundamental
Activity
4 or 32 <Integer> or
<Integer><Integer>
<Integer><Integer>
<Integer><Integer>
<Integer>Integer>
One integer for each sensor
Range 0 - 999
6
(0x06)
Current
Z-Ratio
4 or 32 <Float> or
<Float><Float><Float>
<Float><Float><Float>
<Float><Float>
Z-Ratio currently in use.
7
(0x07)
Anharmonic
Frequency
8 or 64 <Long> or
<Long><Long><Long>
<Long><Long><Long>
<Long><Long>
One long integer for each sensor.
Unconverted, to convert to Hz, multiply by
0.000873114913702011
8
(0x08)
Anharmonic
Activity
4 or 32 <Integer> or
<Integer><Integer>
<Integer><Integer>
<Integer><Integer>
<Integer>Integer>
One Integer for each sensor
Range 0 - 999
9
(0x09)
New crystal
Frequency
16 or 128 <Long><Long> or
< Long>x16
Returns the Fundamental and Anharmonic
"NEW" crystal frequencies
10
(0x1a)
LCQ value 4 or 32 <Float> or
<Float><Float><Float>
<Float><Float><Float>
<Float><Float>
Returns the LCQ running value
(for factory diagnostic use only)
Table 10-14 Status sensor (continued)
SS
Command
ID
Description
Number
of bytes
in the
response
Data Returned Notes

Table of Contents

Related product manuals