10.4.2.3 Allow client access
ü Supervisor rights
1 > Setup > Network > Client access
2 To allow the leak detector to be operated by PC or Tablet, activate the "Client
Access" option.
ð If the option "Client Access" is not activated, the leak detector cannot be
operated via PC or Tablet. No settings can then be changed.
3 Save .
Display of all connected clients
ü The network connection between the leak detector and one or more PCs or Tablets
has been established, see “Setting the wireless LAN connection in the PC or Tablet
[}140]”.
► > Setup > Network > Connected clients