INFICON Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 87 / 152
6.8.8 Call information on the SL3000 Sniffer line
> Sniffer line
6.9 Log
6.9.1 Call the error and warning log
> Logs > Errors and warnings
If there are more than 20 entries always the oldest entry is deleted.
6.9.2 Call calibration log
The entries are for the entire period of use of the device. If there are more than
approx. 20 entries always the oldest entry is deleted.
> Logs > Calibrations
6.9.3 Call result log
For displaying test results in the "Vacuum"(Q
max
) operation mode.
If there are more than 50 entries always the oldest entry is deleted.
> Protocols > Result log
6.9.4 Call maintenance log
> Logs > Maintenance
If there are more than 20 entries always the oldest entry is deleted.
6.10 Device settings
6.10.1 Viewing and customizing individual parameters
You are able to get a quick overview of the actual settings of the device and you can
make modifications, if needed.
ü You have the required right to change the specific parameter.
1 > Parameter sets > Parameter list
ð The device settings will be displayed as a list.
2 Modify individual parameters, if needed.
3 Save .