INFICON Accessories and interfaces | 10
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 135 / 152
Function State: Description
Setpoint 4 closed: Measured leak rate exceeds the setpoint 4
open: Measured leak rate undercuts the setpoint 4
Purge closed: Purge valve open
open: Purge valve closed
Vent closed: Internal vent valve open.
open: Internal vent valve closed
Standby or evacuate closed: State standby or evacuate active
open: State standby of evacuate not active, such
as when measuring
Status calibration
leak valve
closed: Calibration leak valve is open
open: Calibration leak valve closed
Warning closed: Warning
open: No warning
ZERO active closed: ZERO switched on
open: ZERO switched off
10.2.5 Setting up the I/O module protocol
To switch between ASCII and LD protocol.
ü Supervisor rights
1 > Setup > Accessories > I/O module > I/O module protocol
2 Adjust.
3 Save .