INFICON Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 83 / 152
5 To export the displayed cycle, connect a USB flash drive (FAT32 formatted) to
the leak detector and press .
6.7.3 Export measurement data via network
In addition to access to the graphical user interface, you also have the option of
retrieving specific measurement data from your leak detector.
A data interface was implemented for this purpose (REST interface).
This interface responds to requests on port 3000 when valid parameters are
transmitted with data in the requested format.
ü A network connection has been established between the leak detector on one side
and the PC or Tablet on the other side.
See also “Operate leak detector via web browser (LAN) [}137]” or “Operate leak
detector via web browser (wireless LAN) [}139]”.
ü The measurement data were recorded. See also "Switch on data recording and
configure export [}81]".
You have two options for exporting the desired measurement data:
ð You use the convenient options to first display measurement cycles and click
on . See also "Display measurement cycles [}82]".
After activating the from the PC or Tablet, a file with the relevant
measurement data is offered for download.
ð Alternatively, enter your request with the desired parameters as URL in your
browser. With this procedure, you not only set the desired time period (a
block of maximum 24 hours), but you can also select the file format and data
scope (with or without metadata).
When accessing via LAN, use the IP address found under > Device >
Network.
When accessing via WLAN, use the URL found under > Setup > Network
> WLAN.
Example 1:
http://192.168.11.124:3000/md?f=csv_en&c=4&d=mr