4.3.5 Markings on the device
Device cannot be scrapped with the normal domestic waste.
4.4 Technical data
Mechanical data
UL6000 Fab, UL6000 Fab PLUS
Dimensions (L × W × H) 1050 mm×472 mm×1040 mm
Mass approx. 144kg
Inlet flange DN63
Display
Display Color display with capacitive touchscreen
Screen diagonal 7 inch
Resolution 800 x 480 pixel
Colors 16.7 M
Background light LED
Electrical data
UL6000 Fab, UL6000 Fab PLUS
Operating voltage 100 - 240 V ±10%, 50 / 60 Hz
Max. power consumption 1500VA
Typical power consumption
- when pumping
- in measuring mode
1000 VA
800 VA
Max. power consumption 100 V 13A
Max. power consumption 240 V 6A
Short-circuit current rating (SCCR) 300A
Main fuse Circuit breaker 20 A thermal fuse
Protection class EN 60529 IP20
UL 50E Type 1
Overvoltage category II
Electronic interfaces USB, Ethernet, HDMI, SNIFFER,
ACCESSORIES, LD, REMOTE
CONTROL, SERVICE