9 Technical data
82
9.1.2 Thermocouples
Designation Standard Measuring range Measuring accuracy
b
ITS
Fe-CuNi “L” DIN 43710:1985-12 -200 to +900 °C ±0.1 % 68
Fe-CuNi “J” DIN EN 60584-1:2014 -210 to +1200 °C ±0.1 % from -100 °C 90
Cu-CuNi “U” DIN 43710:1985-12 -200 to +600 °C ±0.1 % from -100 °C 68
Cu-CuNi “T” DIN EN 60584-1:2014 -200 to +400 °C ±0.1 % from -150 °C 90
NiCr-Ni “K” DIN EN 60584-1:2014 -200 to +1300 °C ±0.1 % from -50 °C 90
NiCr-CuNi “E” DIN EN 60584-1:2014 -200 to +1000 °C ±0.1 % from -80 °C 90
NiCrSi-NiSi “N” DIN EN 60584-1:2014 -200 to +1300 °C ±0.1 % from -80 °C 90
Pt10Rh-Pt “S” DIN EN 60584-1:2014 -50 to 1768 °C ±0.15 % from -60 °C 90
Pt13Rh-Pt “R” DIN EN 60584-1:2014
Pt30Rh-Pt6Rh “B” DIN EN 60584-1:2014 -50 to 1820 °C ±0.15 % from 400 °C 90
W5Re-W26Re “C” ASTM E230M-11 0 to 2315 °C ±0.15 % 90
W5Re-W20Re “A1” GOST R 8.585-2001 0 to 2500 °C ±0.15 % 90
W3Re-W25Re “D” ASTM E1751M-09 0 to 2315 °C ±0.25 % 90
Chromel®-COPEL® "L" GOST R 8.585-2001 -200 to +800 °C ±0.1 % from -80°C 90
Chromel®-Alumel® GOST R 8.585-2001 -270 to +1372 °C ±0.1 % from -80 °C 90
PLII (Platinel) ASTM E1751M-09 0 to 1395 °C ±0.15 % 90